Objective. Stimulator-attached dissecting instruments are useful for intraoperative nerve monitoring during thyroidectomy. The aim of this study was to evaluate the feasibility of an attachable ring stimulator (ARS) by comparing the electromyography (EMG) amplitudes evoked by an ARS and a conventional stimulator. Methods. Medical records of fourteen patients who underwent thyroidectomy using intraoperative neuromonitoring between June and August 2019 were retrospectively reviewed. The amplitudes of V1, R1, R2, and V2 signals were checked using both the ARS and a conventional stimulator, at the same point. Results. Both stimulators were tested on 20 recurrent laryngeal nerves (RLNs) and 20 vagus nerves (VNs). In all the nerves, the amplitudes of V1, R1, R2, and V2 were greater than 500 mu V. The mean amplitudes of V1, R1, R2, and V2 checked with the ARS were 1175, 1432, 1598, and 1279 mu V, respectively. The mean amplitudes of V1, R1, R2, and V2 checked with the conventional stimulator were 1140, 1425, 1557, and 1217 mu V, respectively. Difference between amplitudes evoked by the two stimulators for V1, R1, R2, and V2 was 77, 110, 102, and 99 mu V, respectively. There was no statistical difference in the amplitudes between the two groups for V1, R1, R2, and V2. Conclusion. The ARS transferred electric stimulation as effectively as the conventional stimulator. It is an effective tool for repeated stimulation and facilitates continuous feedback regarding the functional integrity of nerves during thyroid surgery.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Hoon Yub photo

Kim, Hoon Yub
College of Medicine (Department of Medical Science)
Read more

Altmetrics

Total Views & Downloads

BROWSE