- Polymer-based electronics with low bending stiffnesses and high flexibility, including recently reported macroporous syringe-injectable mesh electronics, have shown substantial promise for chronic studies of neural circuitry in the brains of live animals. A central challenge for exploiting these highly flexible materials for in vivo studies has centered on the development of efficient input/output (I/O) connections to an external interface with high yield, low bonding resistance, and long-term stability. Here we report a new paradigm applied to the challenging case of injectable mesh electronics that exploits the high flexibility of nanoscale thickness two-sided metal I/O pads that can deform and contact standard interface cables in high yield with long-term electrical stability. First, we describe the design and facile fabrication of two-sided metal I/O pads that allow for contact without regard to probe orientation. Second, systematic studies of the contact resistance as a function of I/O pad design and mechanical properties demonstrate the key role of the I/O pad bending stiffness in achieving low-resistance stable contacts. Additionally, computational studies provide design rules for achieving high-yield multiplexed contact interfacing in the case of angular misalignment such that adjacent channels are not shorted. Third, the in vitro measurement of 32-channel mesh electronics probes bonded to interface cables using the direct contact method shows a reproducibly high yield of electrical connectivity. Finally, in vivo experiments with 32-channel mesh electronics probes implanted in live mice demonstrate the chronic stability of the direct contact interface, enabling consistent tracking of single unit neural activity over at least 2 months without a loss of channel recording. The direct contact interfacing methodology paves the way for scalable long-term connections of multiplexed mesh electronics neural probes for neural recording and modulation and moreover could be used to facilitate a scalable interconnection of other flexible electronics in biological studies and therapeutic applications.
- Files in This Item
- There are no files associated with this item.
- Appears in
Collections - College of Science > Department of Physics > 1. Journal Articles
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.